Although FID experiment is useful for setting up NMR instrumentation, it has some disadvantages. The main disadvantage is that the decay constant (or shape) of data acquired with FID is affected not only by the sample properties, but also the homogeneity (or uniformity) of the magnetic field over the sample. Most scientists are only interested in the influence of sample properties on the NMR data, so removing the effect of the magnetic field on the decay of the NMR data would be extremely useful. Fortunately, the effects of the magnet on the signal can be removed by using a special pulse sequence called CPMG sequence.   Fig 1: CPMG sequence CPMG sequence is one of the important means to measure transverse relaxation time of the sample. The CPMG sequence involves applying not a sin...

NMR Sequence—CPMG Sequence

Although FID experiment is useful for setting up NMR instrumentation, it has some disadvantages. The main disadvantage is that the decay constant (or shape) of data acquired with FID is affected not only by the sample properties, but also the homogeneity (or uniformity) of the magnetic field over the sample. Most scientists are only interested in the influence of sample properties on the NMR data, so removing the effect of the magnetic field on the decay of the NMR data would be extremely useful. Fortunately, the effects of the magnet on the signal can be removed by using a special pulse sequence called CPMG sequence.

 

CPMG sequence

Fig 1: CPMG sequence

CPMG sequence is one of the important means to measure transverse relaxation time of the sample. The CPMG sequence involves applying not a single 90 degree RF pulse to the sample, but many 180 degree pulses. Each time a 180 degree pulse is applied, the signal decay to the magnetic field is removed and a single data point is acquired. The decay of the data thus only reflects the sample properties and for liquids is exponential in nature (a distribution). The decay constant of the exponential is called  (the spin-spin relaxation time) or the transverse relaxation time.

CPMG sequence Echo

Fig 2: CPMG sequence Echo

The parameters of this sequence are listed below.

                                      Table 1: The parameters of CPMG sequence

P1

the width of 90-degree RF pulse
DL1 Delay 1, time interval between 90 degree pulse and adjacent180 degree pulse; time interval between 180 degree pulse and adjacent 180 degree pulse is 2*DL1
P2 the width of 180-degree RF pulse
TW the waiting time between two scans of sampling data
EchoCnt Number of 180 degree RF pulse
RFD Time interval between 90 degrees RF and the beginning of signal acquisition
Peaks parity the inversion of sampling points

 

In the tests, the value of τ is set to be 1000μs and only the signal of the first echo, which mainly contains the signal of oil due to the almost complete decaying of fibred signal, needs to be applied. What’s more, the signal of echo is more stable than that acquired by FID sequence and is less influenced by the outside environment. Therefore, this sequence is regarded as the first choice of NMR detection. However, the τ value and echo number will change with different samples.

PQ001

NMR MRI Products

Portable Educational MRI Device

Portable Educational MRI Device

MiniEDU20 MRI system is a portable MRI system,designed for MRI education and training, Including MRI theory,MRI Sequence and MR Imaging.

MRI Contrast Agent Analyzer

MRI Contrast Agent Analyzer

NMI20 integrates industrial design, analysis, and imaging in one easy to use instrument,It enables measurement of T1& T2 relaxation time.

Small Animals MRI Deivce

Small Animals MRI Deivce

MesoMR is designed for in-vivo MRI observation of small animals. It provides high contrast images and has an intuitive user interface.

Rock Core NMR and MRI System

Rock Core NMR and MRI System

The Rock Core NMR and MRI System with High-Temperature High-Pressure System (HTHP), it's suitable for research with temperature and pressure.